Anaesthesia System KAM-B100

Anaesthesia System KAM-B100

Anaesthesia System KAM-B100 is fully featured high-end general anaesthesia workstation equipped with wide screen display for respiratory trends, and other vital parameters. PCV and SIMV + PS modes ensure flexible ventilation. Comprehensive monitoring and intelligent gas flush and cut-off system for critical events improves patient safety.

Download Catalog
Ventilation Modes IPPV, PCV, SIMV + PS, PS and Manual
Tidal Volume 20 ml to 1500 ml
Pressure (Inspiratory) 5 cmH2O – 70 cmH2O
Pressure (Support) 3 cmH2O – 50 cmH2O
Electronic PEEP Off / 3 cmH2O – 30 cmH2O
I: E Ratio 4:1 to 1:8
Display 10.4” TFT Colour
Drive Pneumatically driven and electronically controlled
Pressure Trend Paw, Ppeak, Pplat, Pmean, PEEP
Volume Monitoring MV, VT, Freq
Graph Display P-T, F-T, P-V, F-V
Ventilation Gas (Optional CO<sub>2</sub>, AG, MAC) FiO2
Operational modes Closed, Semi-closed, Semi-open
Gas Supply O2 , Air, N2O
Ventilator Pressure at Inlet 280 kPa ~ 600 kPa
Flowmeter O2 : 0 ~ 10L/min
Air : 0 ~ 12L/min
N2O : 0 ~ 10L/min
Auxiliary O2 : 0 ~ 12L/min
Vaporizer Supports 2 vaporizers
CO<sub>2</sub> Absorption Tank 3 L Autoclavable
Vaporizer Filling methods Pour Fill
Key Fill
Quick-Fil
Anaesthetic Agents Halothane, Enflurane, Isoflurane, Sevoflurane
APL Range Spontaneous breathing (SP) -70 cmH2O
T-slope 0 ~ 2s
Plateau (Inspiratory) Off / 5% – 60% Ti
Trigger Sensitivity 1 – 15 L/min
Alarm High/Low Airway Pressure, Sustained Airway Pressure, High/Low Minute Volume, High/Low FiO2, High Breath Rate, O2 supply failure, Power failure, Apnea alarm, Mute : 110s
Power Supply 100 V ~ 240 V 50/60 Hz
Battery 24 V 4Ah ; 120 min
Dimensions 760 × 1300 × 630mm
Weight 100 Kgs
  • Integrated 10.4” Monitor screen Ventilation modes: IPPV, PCV, SIMV + PS, PS and Manual CO2 Absorption: 3 Litre Autoclavable Tank Flowmeter: minimum 25% O2/N2O mixture concentration Double Vaporizer with Selectatec interlock Comprehensive safety alarms with mute Suitable for central or non-central OT Suitable for Adult, Pediatric and Infant Monitoring of O2, CO2 and anaesthetics (optional) Heated Breathing system (optional)

Used in Surgical or Gynaecological departments to administer appropriate anaesthesia dosage at a specified rate to ensure effective sedation during surgeries and other invasive diagnostic or interventional procedures.

Used in Surgical or Gynaecological departments to administer appropriate anaesthesia dosage at a specified rate to ensure effective sedation during surgeries and other invasive diagnostic or interventional procedures.

C-Arm X-ray Imaging

Mobile C-Arm X-ray KCX-A100

Mobile C-Arm X-ray KCX-A100

X-Ray GeneratorFrequency 40 kHz </br> Power Capacity 3.5 kW'

Tube CurrentX-ray : 20 mA - 63 mA </br> Fluoroscopy : 8 mA'

Catalog

Surgical Diathermy

Surgical Diathermy Machine KSD-A100

Surgical Diathermy Machine KSD-A100

Power Consumption350 W'

Operating Frequency330 kHz ─ 460 kHz'

Catalog
Surgical Diathermy Machine KSD-B100

Surgical Diathermy Machine KSD-B100

Power Consumption300 W'

Operating Frequency416 kHz'

Catalog